7.
MN031-447-S1 EN
MN031-447-S0-20101217-EN
Preparation
Ligation instruments must always be cleaned, disinfected
and sterilized before use. Effective cleaning and disinfection
are essential for effective subsequent sterilization of the
ligation instruments.
1.
Ensure that only adequately validated device- and
product-specific methods are used for cleaning,
disinfection and sterilization and that the validated
parameters are complied with in each cycle.
2.
Observe the applicable national legal regulations and
the hygiene regulations of the hospital or clinic.
The following information on the possible number of
preparation cycles should be regarded as a guideline. The
actual number may vary depending on the stress level.
With a sterilization time of 20 minutes and a sterilization
temperature of 134 °C, the number of preparation cycles of
the individual instrument components is:
Jaw insert A: 20,
Shank tube B: up to 200,
Handle C: up to 200,
HF cable D: up to 300.
100.
NightKNIFE / Ligator
7 Preparation
31