Suitable cleaning agents
Ensure that the selected cleaning agent system fulfils
the following requirements:
The cleaning agent is suitable for the ligation
instrument.
If thermal disinfection is not used, a suitable
disinfection agent with proven effectiveness
(e.g. DGHM or FDA approval or CE mark)
and compatible with the cleaning agents is
also used.
The chemicals that are used are compatible
with the instrument components (see
section 7.4, page 36).
Follow the manufacturer's instructions with regard to the
concentration and exposure time of the cleaning and
disinfection products.
Cleaning and disinfection
NOTE
Risk of damage to the HF cable due to incorrect
placement in the CDM!
Ensure that the HF cable is not kinked or pinched.
1.
Place the instrument components in the CDM.
Ensure that:
the instrument components are positioned to
allow exposure to the rinsing media;
the distal ends of the shank tube and jaw
insert are inserted in the rinsing sleeve;
the drawrod of the jaw insert is slid in so the
jaws are opened as far as possible in the
rinsing tube;
the handle is cleaned in the latched position;
NightKNIFE / Ligator