Bit
Description
14
0
SIOS_CHANNEL_ERROR
1
ZERO_CAL_ERROR
2
SPAN_CAL_ERROR
3
SPAN_CAL_ERROR_COMP
4
SDS_CHECKSUM_ERROR
5
UNDERRANGE_ERROR
6
ERROR_SDS_NOT_SUPPORTED
15
1
CONFIG_CHECK_ERROR
Instructions for use
|
Dräger Polytron
Cause
Electronics error.
Faulty zero calibration.
Faulty span calibration.
Error in sensor electronics.
Error in EC sensor.
Fallen below negative threshold.
Sensor dataset is not supported.
Error in channel configuration.
®
6100 EC WL, Dräger Polytron
Remedy
Contact DrägerService.
Repeat zero calibration.
Repeat span calibration.
Contact DrägerService.
Change the sensor.
Carrying out the zero calibration.
A Change the sensor.
B Contact DrägerService.
A Sensor initialization required.
B Contact DrägerService.
®
Repeater WirelessHART
|
Troubleshooting
®
en
55