Guidance and manufacturer's declaration – electromagnetic immunity
The Otivio AS FlowOx™ unit is intended for use in the electromagnetic environment below. The customer or the end user
of the FlowOx™ should assure that it is used in such an environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of the FlowOx™, including cables,
than the recommended separation distance calculated from the equation applicable to frequency of the transmitter.
Recommended separation distance
�� = 0,6 ∗ √��
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by and electro-magnetic site survey, a should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Immunity test
IEC 60601 Test level
FLOWOX™ 2,0 Patient IFU (610-00013) Rev 9
70