Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Proximity fields from
RF
wireless
communications
equipment
IEC 61000-4-3
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic
site survey should be considered. If the measured field strength in the location in which the FlowOx™ is used
exceeds the applicable RF compliance level above, the FlowOx™ should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating
the FlowOx™.
FLOWOX™ 2,0 Patient IFU (610-00013) Rev 9
3 Vrms
15 kHz to
80 MHz
6 Vrms
15 kHz to 80 MHz in ISM and amateur radio bands
80 % AM at
1 kHz
10 V/m
80 MHz to 2,7 GHz
80 % AM at 1 kHz