9.7 Service menu

Use the service menu to read all important settings
and to check instrument functions. In the code menu,
the service range covers codes 50-57.
Code 50
Display of raw data values of cell
– Actual cell current in µA (Conex
– Actual cell current in mV (Conex
Code 51
Display of calibrated gradient of current (last)
calibration.
®
– In µA/ppm (Conex
®
– In mV/pH (Conex
system selected).
®
Not used with Conex
system selected.
Code 52
Display of calibrated
– asymmetry potential with pH sensor system
– redox offset with redox sensor system
Only used with Conex
Code 53
Display of temperature in °C with temperature
sensor (Pt100) connected
Only used with Conex
Code 54
Display of water sensor depending on selected
type of measuring cell.
– AQC-D1/-D11/-D3/-D13:
Display 100 % = water present
Display 0 % = water missing
– AQC-D2/-D12:
Water sensor speed in %
Display 0 % = 200 rpm
Display 100 % = 800 rpm.
Code 55
Test routine, checking of various instrument
functions:
5500: Test of control relay: relay switches,
flashing display of code. To return to code
level, press [OK].
5501: Test of alarm relay: like 5500
5502: Test of current output: 4 mA are displayed
at the current output, flashing display of
code. To return to code level, press [OK].
5503: Like 5502, except that 20 mA is displayed.
5504: Display test:
If the screen is completely darkened: the
display is OK.
If the screen remains partially bright: the
display is faulty.
®
DIS-D)
®
DIS-PR).
DIS-D)
DIS-PR with pH sensor
DIS-PR with redox sensor
®
DIS-PR!
®
DIS-PR!
Code 56:
Setting of display contrast between 0 % and 100
%.
Code 57:
If a controller has been parameterised, the
current controller performance (y-out) is
displayed in %. Controller stop must not be
activated in the process.
The display is 0.00 % if a controller has not been
parameterised.

9.8 Program version

The program version is displayed when the
instrument is started. Example: V.0.15 28feb2003.

9.9 Reset to default settings

To reset the instrument to the factory (default)
settings, do as follows:
1. Select code 01.
2. Enter factory setting code 6742.
3. Press [OK] to confirm.
All previous settings are reset.
See sections
6.5 Factory settings
overview.
and
9.4 Code
23