± 0.5 in. (± 1.3 cm)
6. Center the mounting plate (D) on the mark that was made in step 5 and use a pencil to mark the bottom of the mounting
plate (D) (Figure 3).
N N o o t t e e - - The bottom of the mounting plate must be 4 in. - 5 in. (10.2 cm - 12.7 cm) from the floor.
7. Using a level on the bottom of the mounting plate (D), make sure that the bottom of the mounting plate (D) is level when
you position the mounting plate (D) on the reference marks made in steps 5 and 6 (Figure 3).
8. Use a pencil to mark the three screw holes of the mounting plate (D).
9. Using the appropriate tool with the hospital supplied fasteners (E, not included), secure the mounting plate (D) to the
wall (Figure 3).
10. Attach the S S e e c c u u r r e e C C o o n n n n e e c c t t (F) to the mounting plate (D) (Figure 3).
11. Using a #2 Phillips screwdriver, secure the S S e e c c u u r r e e C C o o n n n n e e c c t t (F) to the mounting plate (D) with the two supplied screws
(700001126359) (G) (Figure 3).
12. Plug the S S e e c c u u r r e e C C o o n n n n e e c c t t power supply into a hospital grade protective earthed wall outlet.
N N o o t t e e - - Position the power supply in an accessible location.
13. On the S S e e c c u u r r e e C C o o n n n n e e c c t t , plug in the female end of the power supply.
14. On the S S e e c c u u r r e e C C o o n n n n e e c c t t , plug in the nurse call communication cable.
15. Using a #1 Phillips screwdriver, secure the nurse call communication cable to the S S e e c c u u r r e e C C o o n n n n e e c c t t .
16. Connect and secure the nurse call communication cable to the nurse call system wall plug.
17. Configure the S S e e c c u u r r e e C C o o n n n n e e c c t t , see Configuring the S S e e c c u u r r e e C C o o n n n n e e c c t t (page 10).
18. Follow the procedure in the product manual to connect the product to the S S e e c c u u r r e e C C o o n n n n e e c c t t .
C C A A U U T T I I O O N N - - Always match the dip-switches of a product to the head wall configuration if a nurse call communication
cable needs to be connected to avoid the risk of head wall damage.
N N o o t t e e
• If a S S e e c c u u r r e e C C o o n n n n e e c c t t is moved, repeat steps 1 and 10-17.
• If a product is moved to another configured S S e e c c u u r r e e C C o o n n n n e e c c t t , no change needs to be made as the product will
connect automatically.
5212-009-205 Rev AA.1
F F i i g g u u r r e e 3 3 – – M M o o u u n n t t i i n n g g s s p p e e c c i i f f i i c c a a t t i i o o n n s s
4 in. - 5 in. (10.2 cm - 12.7 cm)