11 9.6 Dimensions Eps - Bosch EPS 815 Manual Original

Ocultar thumbs Ver también para EPS 815:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

9.6
Dimensions EPS 815
a
b
200mm 200mm
1 Maximum installation space taken up by measuring cell carrier of
KMA 802/822 or measuring-glas carier of MGT 812/824
2 Maximum installation space taken up by monitor and keyboard of
KMA 802 or stroge and rev counters of MGT 812/824
3 Maximum installation space taken up by PC housing of KMA 802
4 Power connection
5 Water connection (Coolant connection)
a) Supply connection G½" A ISO 228- 1
b) Return-line connection G½" A ISO 228- 1
!
The operator must provide a pressure reducer (set
to 250 kPa), a safety filter and a shut-off valve in the
coolant supply.
!
The coolant temperature must not exceed 17 de-
grees. If the coolant temperature is higher than this,
the operator must provide a suitable cooling device.
!
The coolant must be discharged via an oil separator into
the waste water lead. No shut-off valves must be used.
Robert Bosch GmbH
1
2
458753 R
4
2306 mm
2937 mm
Technical Data | EPS 815 | 23
3
4
5
673 mm
310 mm
1 689 989 348
en
|
2017-07-01

Publicidad

Capítulos

Tabla de contenido
loading