Automatic Cleaning
If you wish to use automatic cleaning for these devices, you must follow the washer
manufacturer's recommendations pertaining to these types of devices.
1. Ensure all pre-processing instructions are followed prior to cleaning.
2. Clean the sterilization containment systems via the automatic cleaning
parameters below.
3. Manipulate the sterilization containment systems to allow rinse water to drain.
MINIMUM
PHASE
RECIRCULATION TIME
Pre-Wash
15 Seconds
Enzyme Wash
1 Minute
Wash
2 Minutes
Rinse
15 Seconds
Pure Rinse
15 Seconds
Drying
N/A
4. If visible moisture is present, dry the instrument with a clean, lint-free towel.
5. Visually examine each containment system for cleanliness.
6. If visible soil remains, repeat cleaning procedure.
Note: Do not use ultrasonic cleaning for anodized aluminum containment system.
Inspection/Maintenance
Proper care and handling is essential for satisfactory performance of any surgical
device. The previous cautions should be taken to ensure long and trouble-free
service from all your surgical devices. Inspect devices before each use for broken,
cracked, tarnished surfaces, movement of hinges, and chipped or worn parts. If any
of these conditions appear, do not use the device. Return devices to an authorized
repair service center for repair or replacement.
Before sterilizing, lubricate the device with instrument milk or a steam permeable/
water soluble lubricant, following the lubricant manufacturer's instructions.
Let devices drip dry for three (3) minutes before packaging for sterilization.
Packaging
Sterilization wrap material must be cleared for the applicable sterilization
modality by your country's regulatory body. Use in accordance with packaging
manufacturer's sterilization instructions being sure to protect jaws and cutting
edges from damage.
Prevacuum Steam Sterilization Parameters
Preconditioning Pulses: 3
Temperature: 132°C (270°F)
Exposure Time: 4 minutes
Dry Time: 30 minutes in chamber; 15 minutes chamber door cracked and
60 minutes on drying rack (may vary according to load contents)
Sterilization Configuration: Wrapped (2 layer 1-ply or 1 layer 2-ply)
Total Weight: 6.82 kg (15 lbs)
DETERGENT TYPE AND
WATER
CONCENTRATION
TEMPERATURE
(IF APPLICABLE)
Cold Drinking /
Potable Water
N/A
1°C - 16°C
(33°F - 60°F)
• Detergent: pH-neutral
Hot Drinking /
/enzymatic
Potable Water
• Concentration: Per the
43°C - 82°C
detergent manufacturer's
(110°F - 179°F)
recommendations
• Detergent: pH-neutral
Hot Drinking /
/enzymatic
Potable Water
• Concentration: Per the
43°C - 82°C
detergent manufacturer's
(110°F - 179°F)
recommendations
Hot Drinking /
Potable Water
N/A
43°C - 82°C
(110°F - 179°F)
Purified Water
43°C - 82°C
N/A
(110°F - 179°F)
N/A
N/A
Sterilization for outside United States market
Prevacuum Steam Sterilization Parameters
Minimum Preconditioning Pulses: 3
Minimum Temperature: 132°C (270°F)
Minimum Exposure Time: 3 minutes
Minimum Dry Time: 30 minutes in chamber; 15 minutes chamber door cracked
and 60 minutes on drying rack (may vary according to load contents)
Sterilization Configuration: Wrapped (2 layer 1-ply or 1 layer 2-ply)
Total Weight: 6.82 kg (15 lbs)
Gravity Steam Sterilization Parameters
Temperature: 132°C (270°F)
Exposure Time: 40 minutes
Dry Time: 30 minutes
Sterilization Configuration: Wrapped (2 layer 1-ply or 1 layer 2-ply)
Total Weight: 6.82 kg (15 lbs)
100% Ethylene Oxide (EO) Sterilization Cycle
EO Sterilant Concentration: 100% EO, 725mg/L
Preconditioning Time: 30 minutes
Temperature: 55°C (130°F)
Exposure: 60 minutes
Relative Humidity: 50-80%
Aeration: 8 hours @ 43°C
Sterilization Configuration: Wrapped (2 layer 1-ply or 1 layer 2-ply)
Total Weight: 6.82 kg (15 lbs)
Storage
After sterilization, devices must remain in sterilization packaging and be stored in a
clean, dry environment.
Cooling
• Do not remove sterilization containment systems from the carrier until cool to
touch or place warm containment systems on cold tabletops. Proper cool down
is necessary to prevent wet sets or re-condensation from forming.
• Eliminate drastic temperature differences by keeping the sterilization
containment systems away from cool ventilation ducts or cool drafts. Rapid
cooling can cause wet sets.
• Do not place processed sterilization containment systems into the sterile storage
area until they reach room temperature.
Warranty
CareFusion offers a lifetime guarantee on every surgical device bearing the
V. Mueller brand name (unless otherwise noted) to be free of functional defects in
workmanship and materials when used normally for its intended surgical purpose.
Any V. Mueller device proving to be defective will be replaced or repaired at no
charge.
Repairs or modifications performed by unauthorized personnel may void all product
warranties and could affect performance and efficacy of device.
Repair Service
Regardless of age, if any V. Mueller device needs service, return it to an authorized
repair service center. For repairs outside the U.S., please contact your local
distributor.
Note: All devices being returned for maintenance, repair, etc. must be cleaned and
sterilized per these instructions for use prior to shipment.
Contact Information
BD Customer Service
800-323-9088
For email inquiries:
Customer_Support@BD.com
www.bd.com
For customers outside of the USA, please contact your local distributor.
Other Resources: To learn more about sterilization practices and what is required of
manufacturers and end users, visit www.aami.org, www.aorn.org or www.iso.org.