2.3 Front Panel

1. Power button
2. Pressure Adjustment
The function can automatically adjust the mattress pressure based on the patient's weight. To
remind user the process is on-going, the function LED lights will periodically flash and be
consistanly on after the process is completed. There are three conditions will trigger this function
to ensure the appropriate pressure setting for the patient.
A) When in therapy mode, press the TUNING button for 3 seconds, system will automatically start
the readjustment of the pressure setting.
B) The system will automatically start Auto-Adjust immediately after the initial mattress inflation is
completed.
C) While the system detects a significant change in patient's weight on the mattress (e.g. patient
Ingress) for a period of time, the system will automatically run the function again.
NOTE: Every time the mattress is initialized (inflated), it will automatically go to "Max Firm"
mode to hasten inflation. Once the system is ready to use, the system will go to
alternate mode automatically.
NOTE: You can also check if the pressure is suitable for the patient by sliding one hand
beneath the air cells at the level of the patient's buttocks. Always leave at least 1"
inch space between patient and the static cell to prevent bottoming out.
NOTE: During the Auto-Adjust operation, it is normal that system will go through a series of
inflation and deflation.