HBS x-touch 200i

6 Accessories

In order to avoid incompatibility and unsafe operation, only use the accessories recommended by the
manufacturer when working with the HRF 100 x-touch.
6.1

Accessories for the monopolar operating mode

Handles for monopolar electrodes can be connected to the unit using a 2-pin standard or MARTIN
plug or using the 4 mm plug (banana plug) with only one contact.
3-pin standard plug
Neutral electrode
The HRF 100 x-touch detects neutral electrodes with undivided and divided contact surface. A
comparison of the electrode detected by the machine and the used electrode provides a direct
indication of correct adhesion of the neutral electrode. Both single-use and reusable neutral
electrodes can be used.
Request a copy of our extensive
electrosurgery accessories catalogue today!
MARTIN monopolar
Valleylab plug for the neutral electrode
HEBUmedical recommends the use of the single-use safety
neutral electrode with double divided contact surface
(HF9564H) for maximum safety!
plug
EN
Monopolare 4mm 1-pin plug
(banana plug)
43