Technical Data - HEBU medical HRF 100 x-touch Instrucciones Para El Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

EN

5 Technical data

Mains connection
Supply voltage
Power consumption
Continuous operation
Nominal frequency
Output power in monopolar operation
Pure cut
Blend cut
Soft coagulation
Forced coagulation
Output power in bipolar operation
Bipolar coagulation (Mode „Bipolar Forceps")
Bipolare Koagulation (Modus „Bipolare Pinzette
AUTO Start")
AUTO DELAY
AUTO STOP
Safety measures
Type
Protection class
Equipotential connection
Safety
Electromagnetic compatibility
Safety of high-frequency surgical units
Miscellaneous
Operating temperature
Humidity during operation
Air pressure during operating
Storage temperature
Storage humidity
Air pressure in storage
Dimensions (L/W/H)
Weight
42
100-260 Volt, 50 - 60 Hz.
160 VA (100 Volt, PF = 1,0)
150 VA (230 Volt, PF = 0,99)
Interval 10s / 30 s (activation / pause)
500 kHz
100 W at 400 Ω
100 W at 400 Ω
100 W at 50 Ω
100 W at 400 Ω
100 W at 50 Ω
100 W at 50 Ω
Yes
Yes
CF
I
Yes
In accordance with DIN EN ISO 60601-1
In accordance with DIN EN ISO 60601-1-2
In accordance with DIN EN ISO 60601-2-2
+10°C to +40°C
30-75%
700-1060hPa
-10°C to +60°C
10-85%
500-1060hPa
380 x 370 x 145 mm
7,0 kg
HBS x-touch 200i