Section 2 Installation
GF100 Series
2-17 Introduction to the MultiFlo

2-18 Using the MultiFlo Configurator

2-10
TM
Configurator
The MutliFlo Configurator application is used to configure the gas and
range of the GF Series devices. The following section describes the
MultiFlo Configurator and its uses.
The MultiFlo Configurator application allows communication to GF Series
devices through personal computer with serial COM Port and a Windows
10, Windows 7 or Windows XP operating system. It's primary function is to
configure gas and flow ranges within six defined standard configurations.
Flow ranges are configured to the Nitrogen equivalent.
Using the MultiFlo Configurator software, configure the gas and flow rate
according to Table 2-3.
Table 2-3 Gas and Flow Ranges - MultiFlo Configurable - N2 Equivalent
Standard MG-MR Bin
Configurations
SH40
SH41
SH42
SH43
SH44
SH45
SH46
SH47
SH48
SH49
SH50
The MultiFlo Configurator interfaces to the GF Series device through
RS485 or DeviceNet communications. There are various ways to connect
the device regardless of device configuration. Devices may be connected
through the diagnostic port using cables in one of the two Basic MultiFlo
Configurator Kits or DeviceNet devices can alternatively be connected
using a National Instruments or SST DeviceNet scanner card.
778Z010ZZZ Basic MultiFlo Configurator Kit
A331710003 Cable Assembly 2.5mm
214F027AAA USB-RS485 Converter with DB-9 female
778Z011ZZZ Basic MultiFlo Configurator Kit
w/Power Supply and Adapter Cables
A331710003 Cable Assembly 2.5mm
214F027AAA USB-RS485 Converter with DB-9 female
A332295001 Power Supply MFC
A332297002 Cable, Power, 9-Pin
A332297001 Cable, Power, DeviceNet
Connect the MultiFlo Cable Adapter 2.5mm jack to the Diagnostic Port on
the top of the device. See Figure 2-12.
Connect the RS485 end of the converter to the 9-Pin RS485 end of the
MultiFlo Cable Adapter.
Installation and Operation Manual
X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
Flow range
Gas Flow Range
Code
(N2 Equivalent)
010C
3-10 sccm
030C
11-30 sccm
092C
31-92 sccm
280C
93-280 sccm
860C
281-860 sccm
2.6L
861-2600 sccm
7.2L
2601-7200 sccm
015L
7201-15000 sccm
030L
15001-30000 sccm
040L
30001-40000 sccm
055L
40001-55000 sccm
April, 2017