Display
Access
PV
READ
PV
READ
PV
READ
PV
READ
DIAG_STATUS
READ
DIAG_STATUS
READ
DIAG_STATUS
READ
DIAG_STATUS
READ
For further information see: "Troubleshooting", page 46.
8.4

Repeater feature

The feature can be enabled or disabled as part of the network integration. For
gas detector energy optimization, disabling the feature is recommended when
there is no need for it due to the network topology.
9

Operation

9.1

Special states

In a special state, correct measurement or alarm activation is not guaranteed.
Special states are as follows:
Gas concentration outside measuring range
Error
Calibration
Bump test
Warm-up phase
Maintenance mode
Instructions for use
|
Dräger Polytron
Byte no.
Content
2
PV_Value (1/4)
3
PV_Value (2/4)
4
PV_Value (3/4)
5
PV_Value (4/4)
1
Diag_Status (1/4)
2
Diag_Status (2/4)
3
Diag_Status (3/4)
4
Diag_Status (4/4)
®
6100 EC WL, Dräger Polytron
Modbus input register
Register N+1
Register N+2
Register M
Register M+1
Special states are displayed in PolySoft.
9.2

Exiting special states

1. Identify the malfunction:
a. Check measured value status (PV_STATUS).
b. Check device status (DIAG_STATUS).
c. Read out the device status using PolySoft.
2. Start with fault elimination (see "Troubleshooting", page 46) or follow
corrective measures according to PolySoft.
Errors and warnings are non-latching. If error or warning conditions are
resolved, the respective special state is also no longer transmitted.
10
Calibration
The configuration is done with PolySoft.
®
Repeater ISA
|
en
Operation
Data type
32-bit float (4 bytes),
Float32 acc. to IEEE 754
UINT (16 bit)
UINT (16 bit)