Commissioning
7
Commissioning
1. Turn on the power supply or install the battery box (see "Installing the
battery box", page 41).
 Start sequence begins.
Green LED flashes.
 LED test in progress.
Yellow and red LEDs flash (repeater: Yellow only).
 Warm-up phase 1 starts.
Measured value status is "Bad" (substatus "warm-up phase 1") and
yellow LED flashes, blue LED flashes (Bluetooth
 The start sequence is complete and the gas detector is ready for
operation.
 Warning signal is output.
Green and yellow LEDs flash.
2. Integrate the gas detector into the network (see "Network integration",
page 42).
3. Wait for warm-up phase 1 and 2.
The duration of the warm-up phase depends on the sensor used (see
sensor data sheet).
Settings with PolySoft can already be made.
 At the end of warm-up phase 1, the measured value is available, the
measured value status is "Good" and the yellow LED is off.
4. Calibrate the gas detector (see "Calibration", page 43) as soon as warm-up
phase 2 is completed.
8.3
ISA100 Wireless
The gas detector outputs ISA100 Wireless
Display
Access
PV
READ
42
®
is available).
standard objects
standard objects with attributes for process values (PV) and diagnosis status (DIAG_STATUS) and PROFIsafe data.
Byte no.
1
Instructions for use
5. Test the signal chain by applying target gas to the sensor.
8
Network integration
8.1
Network integration
For network integration, also pay attention to the documentation of the Access
Point and System Manager used.
For a Yokogawa infrastructure, the creation of a YPIF file (consignment file) is
supported by PolySoft.
See technical manual "Wireless Communication ISA100-Wireless
91 00 000, download at www.draeger.com/ifu) for information on:
Recommended network components
Creating the provisioning file (see also the PolySoftHTML help pages)
8.2

Signal transmission to the controller

Publication Rate: 1x every 10 seconds (Yokogawa Management Station)
Measured value updates (publication rate) and timeout times can be set via
the Yokogawa infrastructure.
If there is no sample gas, the gas detector only outputs data every other time
window (power-save mode). If sample gas is present in sufficient concentration,
all time windows are used.
Content
Modbus input register
PV_Status
Register N
|
®
Dräger Polytron
6100 EC WL, Dräger Polytron
TM
" (order no.
Data type
UINT (16 bit)
®
Repeater ISA

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Polytron repeater isa

Tabla de contenido