OPERATION
sEMG TRACTION TREATMENT CONTINUED
Press the sEMG Traction button from
the Home Screen.
NOTE: Once the sEMG Traction
button is pressed, a
continuous audible tone will
sound from the Patient
Interrupt Switch to indicate
sEMG Traction activity
readings.
Set the parameters for sEMG Traction
treatment, using the following
buttons on the Treatment Screen:
View Graph
The View Graph button allows the
clinician to toggle between the
Treatment Graph to the sEMG graph.
sEMG Graph
The sEMG Graph allows the setting
target level of muscle activity for
triggering traction therapy to start.
There are three buttons and a scale to
set the sEMG target:
38
Triton® Traction
Target button - This toggle button
allows the clinician to select the
target either manually or take the
average of muscle activity achieved
over a 15 second period.
Adjust/Capture Target button -
Once the target is selected, this
button allows the clinician to either
adjust the manual target or capture
the average target.
To Adjust the Manual Target -
Press the Target button until Manual
is selected, press the Adjust Target
button, press the Up or Down Arrow
buttons to adjust the Target to the
prescribed level, and press the Return
Arrow to accept. The target range is
from 5 to 2,000 μV.