Troubleshooting
Error
Liquid droplets appear in
the Trough
Tube is blocked during a
run.
Program is automatically
aborted and the
instrument displays the
message "Motor is blocked.
Program was aborted."
Tube comes off the
instrument
A warning message
appears in a pop-up
window when saving an
edited program. "Input
error - Maximum rounds
per run (3200) exceeded!"
New program containing
iterations cannot be saved.
The instrument displays an
error message with the
cause of the error.
A warning message "The
heater was removed!"
appears in a pop-up
window, while running a
heaters program.
Cause of error
Tube is not closed properly.
Sample is jammed between the cap and the
tube and damages the seal. This might cause
liquid to run trough the seal.
Tube seal is leaky.
Sleeve is worn.
Inappropriate sample material.
If the motor is blocked due to a high
resistance of the sample material, the
instrument will try three times to get started
again.
If the blockage can be resolved, the
instrument will continue the run of the
program. If the blockage cannot be resolved,
the program will automatically be aborted.
Tube is not correctly fitted into the snap-fit of
the Trough.
Tube is not closed properly.
The duration of the program exceed 3,200
rounds per run (rpr). It is not recommended
to run programs exceeding this limitation!
Miltenyi Biotec does not guarantee proper
function of the instruments and disposables
above this value.
"Loop command is faulty. End loop command
is missing." indicates that the end loop
command has not been specified.
"Loop command is faulty. The command is
nested." indicates that there are more than
one loop commands.
"Loop command is faulty. Loop has no body."
indicates that there is no command between
the loop and end loop commands.
The heater was removed.
The heater is not installed properly.
Remedy
Close the tube tightly beyond
first resistance.
The top area of the cap and
the tube must be perfectly
clean for proper functioning of
the seal.
Replace the tube.
Exchange Sleeve.
Refer to the respective data
sheet for technical
specifications of the tubes.
Process a smaller amount of
tissue, or use a different
gentleMACS Program.
Check, if tissue is trapped
between the stator and rotor.
Release tissue, reinstall tube,
and press continue to resume
the program.
Ensure correct installation of
the tube.
Close the tube tightly beyond
first resistance.
Press OK to save the program.
Press Abort to re-edit the
program.
Close iterations by adding an
end loop command.
Use only one loop command
and close the iteration by
adding an end loop command.
At least one program step
must be included between a
loop and end loop command.
Add steps and define
commands that shall be
iterated.
Reinstall heater.
Make sure that the heater is
installed properly.