Technical data
Maximum noise emissions
Sound pressure level at workstation
Sound power level
Measurement operation conditions:
8x C Tubes containing 500 µl of liquid,
processed at 4,000 rpm
Working temperature heaters
Table 7.1: Technical specifications of the gentleMACS Octo Dissociator with Heaters.
The gentleMACS™ Octo Dissociator with Heaters is a Protection Class I instrument and must be plugged into a
ground power outlet. The main power supply cord and plug of the instrument must comply with following
specifications (USA and Canada only): UL listed and KAM cord, minimum type SJ, minimum 18 AWG,
3 conductors. Rated for a minimum temperature of 60 °C. Provided with grounding-type (NEMA 5-15P-9
attachment plug, rated 125 VAC, 10 A. Opposite end terminates in IEC 320 style connector, rated 125VAC, 10A.
Conditions of operations: Supply voltage fluctuations up to +/- 10% of the nominal voltage. Transient over-
voltages present on the main supply: category II.
The instrument is suitable for rated pollution degree 2.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a particular installation.
7
If this equipment does cause harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
Reorient or relocate the receiving antenna.
l
Increase the distance between the equipment and the receiver.
l
Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
l
Consult the dealer or an experienced radio/TV technician for help.
l
Changes or modifications of the instrument, unless expressly approved by Miltenyi Biotec, may void authority to
operate the instrument pursuant to FCC 47 CFR.
For other safety considerations, refer to the product label, or visit
Design and specifications are subject to change without notice.
82
Specification
86 dB(A) +/- 2.5 dB(A)
92 dB
37 °C
www.miltenyibiotec.com .