22
Calibrating and Taking a Reading
1. From the Home screen, select TcB Reading.
Note: If you set Nurse ID and Patient ID to None in the Setup menu, Steps 2
and 3 are not required.
2. You can enter a Nurse ID and two of the following forms of
Patient ID:
–
Patient Name
–
Patient Medical Record Number
–
Patient Date of Birth
3. To enter information, scan the appropriate barcode or type the
information using the on-screen keyboard. To scan a barcode:
a. Highlight Scan In to activate the barcode scanner. Point
the scanner on the hand-held unit towards the barcode and
press the
b. The scanned information will appear on the screen. If it is
correct, select Confirm by pressing the
To type in the information:
a. Using the on-screen keyboard, type in the appropriate
Nurse and/or Patient ID. Use the arrow keys to move to
the appropriate position on the keyboard and press the
(Enter) key to select each letter/number.
b. Highlight Done and press the
settings.
BiliChek user manual
(Enter) key.
(Enter) key.
(Enter) key to save the
TcB Reading Selection
(
Note: Press the
Help) key
on the BiliChek keypad for more
information during any of these
steps.
Note: This unit is equipped with
a Critical Power-off feature. After
completing the calibration, the unit
will beep (unless the sound is set
to level 0) if no measurement is
performed within a preset period
of time. The unit will power-
off 2 minutes later if no key is
pressed. If the unit automatically
shuts off, it is necessary to
perform another calibration after
restarting.
Note: To exit during calibration
or measurement, press the
(Home) key to return to the home
screen. Data will not be saved.