R82 Manatee Manual Del Usuario página 20

Tabla de contenido

Publicidad

Safety
Signs, symbols and instructions placed on this
product must never be covered or removed and
must remain present and clearly legible throughout
the entire lifetime of the product. Immediately
replace or repair illegible or damaged signs, symbols
and instructions. Contact your local dealer for
instructions.
EN
Caregiver
Read the "User Guide" carefully before use
and keep it for future reference. Incorrect use
of this product may cause serious injury to the
user
Always use the correct lifting techniques and
aids for this purpose
Never leave the user unattended in this
product. Ensure permanent supervision by an
adult
Repair/replacements must only be made using
new original R82 spare parts and fittings and
carried out in accordance with the guidelines
and service intervals prescribed by the supplier
If there is any doubt as to the continued safe
use of your R82 product or if any parts should
fail, stop using the product immediately and
contact your local dealer as soon as possible
Environment
Observe the surface temperature on the
product, before placing an user in it. This goes
especially for users with insensitive skin as they
cannot feel the heat. If the surface temperature
is above 41°C then allow the product to cool
before use
User
If the user is close to the max. load limit and/
or has a lot of involuntary movements, e.g.
rocking, a larger size with a higher max. load or
a different R82 product should be considered
The head of a user must always be within the
top of the frame
Product
Carry out all positional adjustments on the
product and accessories and ensure that
all knobs, screws and buckles are securely
fastened before use. Keep all tools out of the
reach of children
Ensure the stability of the product is
maintained before placing the user in it
Please check on a regular basis that the
zippers are safely closed.
Check that all release buttons mesh into
place after adjustment
Ensure the Anti-slip rubber covers are
functional
Check on the cover, belts and accessories
and make sure that the Velcro fastens
securely
The enclosed belts and accessories cannot
be used as long-term supports but may only
be used as an extra help when positioning
the user.
Service information
Maintenance of a medical device is wholly the
responsibility of the owner of that device. Failure
to maintain a device in accordance with the
instructions may invalidate the device's warranty.
Furthermore, failure to maintain a device may
compromise the clinical condition or safety of users
and/or their caregivers.
If the product is subject to tap water containing
chlorine, then discoloration of plastics and fabric
can be expected.
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
After every use
Rinse the product with water to remove
leftovers from soap and organic material
Daily
Clean the anti-slip rubber covers, using a cloth
with warm water and mild detergent and wipe
them dry
Visual inspection. Repair or replace any
damaged or worn-out parts
Check all fittings and buckles for signs of worn
out parts
Ensure that all fasteners are in place and fitted
correctly
Be aware of any MAX mark indications before
adjustment
Monthly
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Every 6 months
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substandard
or faulty
20