EasyMax
102 Quick Start Guide
This document will guide you through the steps of experiment setup,
changing parameters in real time, data visualization and data export on
the EasyMax
102.
For in-depth information on installation and use of the EasyMax
please refer to the installation guide and the operating instructions.
1. Experiment Setup
Insert the reactor into the thermostat.
You have the choice of magnetic or mechanical stirring.
If you select magnetic stirring, place the magnetic bar
into the reactor.
Fill in starting materials, solvents, reagents, etc.
Secure the cover of the 100-mL reactor with the chain
clamp. Make sure the Tr (temperature of the reactor
contents) sensor is immersed at least 1.5 cm into the
reaction mass to provide correct data.
When working with one zone only, you may still want
to plug in the Tr sensor for the unused zone to avoid an
error message for that zone.
Further information can be found in the Operating
Instructions.
2
Heating in Tr Mode
102,
Heating in Tj Mode
In the System Overview you can monitor and change for both zones:
Tr
Tr-Tj – heat flow trending
Tj
R
2. Running an Experiment
Switch on the EasyMax
102 by pressing the push button on
the front panel.
After startup, the touchscreen displays the System Overview.
Zone 1 View
Reflux Mode
Stirrer Speed
– temperature of the reactor contents
– jacket temperature
– stirrer speed
Starting an experiment is done in Single Zone View.
Switch to Single Zone View by touching either the Zone 1
View or Zone 2 View buttons (see above).
The Experiment button lets you name, start, and stop an
EasyMax
102 experiment.
Stirrer
Touching the R button opens a numeric keypad for entering
the desired stirrer speed. Mechanical stirring is the default
whenever a stirrer is plugged in, otherwise magnetic stirring
is activated.
Zone 2 View
Stirrer On/Off
Light On/Off
System Overview