Setting up custom obstetric tables
You must enable all macros in Excel when you use the OB Custom Table Editor in Windows.
When you are finished using the OB Custom Table Editor, you can revert the macro settings in Excel if
desired. For instructions, see Excel Help.
Worksheets & reports
The gender selection from the patient information form is not always correctly reflected in
non-English worksheets. In Spanish, the gender is switched. In German, French, and Italian, the
gender may be blank. When you end the exam, the read-only patient report displays the correct
gender.
DVR recording
To record to the DVR
Wait at least five seconds between removing and reinserting a USB memory stick into the DVR USB
port.
To copy DVR recordings to another device
If video recorded from the DVR is too bright when you view it on a personal computer, adjust the
video color settings on your PC's monitor. For example, change the Dynamic Range to Full. See your
computer's Help for information about available video color settings.
Clinical monitor
WARNING: FUJIFILM SonoSite recommends against using a monitor other than the clinical monitor
provided by FUJIFILM SonoSite. Using a different monitor may result in image distortion and
degradation of image quality.
Scanning
Display formats
When you change display formats (for example, from a single image to Dual, or by changing the
Display Format), the images may overlap. If the M line or D line is located near the side of the image,
it may be hidden. Toggle the trace to expose the M line or D line.
Dual
When you adjust an image in frozen Dual mode, the image may appear incorrectly. To resolve this
issue, pan the image or drag the cine buffer slider.
ECG
If you defibrillate a patient while the ECG module is connected to the system, the ECG signal may
display incorrectly, and the ECG module may need to be replaced.