9.3

Dismantling

1. Disconnect the APC probe from the connecting cable.
9.4

Required aids

Processing step
Items Needed

Precleaning

Soft plastic brush/soft cloth
50 ml disposable syringe with male
Luer-Lock connection
Erbe irrigation adapter 20132-203
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
10 ml disposable syringe with male
Luer-Lock connection (4 ea.)
Erbe irrigation adapter 20132-203
Compressed air gun (2 bar maximum)
Machine cleaning/
Erbe irrigation adapter 20132-203
disinfection
9.5
Precleaning
Preclean the product up to a maximum of 2 hours after use.
For precleaning, you will need water, a 3 % hydrogen peroxide
solution, as well as a mildly alkaline (pH value between 7 and 10)
cleaning agent suitable for ultrasonic use with medical devices
comprised of plastic and metal.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.