Connecting the product
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the adapter.
IMPORTANT! The adapter is only suitable for connecting argon
plasma applicators. APC probes must not be connected. If APC
probes are connected to the adapter, however, a message ap-
pears on the device which must be observed.
1. Switch on the VIO electrosurgical unit and select the basic
program.
2
2. Insert the HF connector of the adapter (1) into the multifunc-
tional receptacle of the APC 2. The connector must click into
place audibly. A message appears on the unit that can be
skipped and that disappears as soon as an argon plasma ap-
plicator is connected to the adapter.
3. Screw the argon gas connection of the adapter (2) into the
argon gas receptacle of the APC 2.
4. When the adapter is connected, a standard setting for mode,
power, effect and gas flow is automatically made on the VIO
electrosurgical unit. However, the settings can be modified
within a certain range if required.