Disconnecting The Product From The Unit; Cleaning, Disinfection, Sterilization; Safety Instructions - Erbe 20132-249 Manual Del Usuario

Adaptador apcapplicator para apc 2 con enchufe apc
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
OBJ_DOKU-194452-003.fm Seite 23 Donnerstag, 2. April 2020 2:47 14
4.4

Disconnecting the product from the unit

1. Lateral forces can damage the receptacle. Therefore ensure
you pull the HF connector of the adapter out of the multi-
functional receptacle in a straight line.
2. Unscrew the argon gas connection of the adapter and pull it
out of the argon gas receptacle in a straight line.
5

Cleaning, disinfection, sterilization

5.1

Safety instructions

Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
23

Publicidad

Tabla de contenido
loading

Tabla de contenido