5.2

Reprocessing limitation

Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
5.3

Dismantling

1. Detach the adapter from the argon plasma applicator. The
argon plasma applicator cannot be reprocessed.
5.4

Required aids

Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
50 ml disposable syringe with male
Luer-Lock connection
Female-to-female luer-lock adapter
(=irrigation adapter)
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
50 ml disposable syringe with male
Luer-Lock connection (4 ea.)
Female-to-female luer-lock adapter
(=irrigation adapter)
Compressed air gun (2 bar maximum)
Machine cleaning/
Female-to-female luer-lock adapter
disinfection
(=irrigation adapter)
24