Intended use

The bipolar LAP forceps are intended for coagulation in the case
of laparoscopic procedures.
The bipolar LAP scissors are intended for the mechanical sepa-
ration of tissue with simultaneous bipolar coagulation of the cut
edges in the case of laparoscopic procedures.
3

Normal use

This product is developed for connection to the bipolar output of
HF generators for use with COAG modes. A power output of 60 W
in bipolar mode must not be exceeded.
The product is used with the following devices:
• Erbe ICC
• Erbe VIO
With Erbe VIO the BIPOLAR FORCED COAG mode must not be
used because the instrument can be damaged due to the high
peak voltage.
4

Maximum electrical capacity

The maximum electrical capacity is given on the instrument in
[V
].
p
5

Safety instructions

WARNING!
This product may be used only by trained medical staff, who
have been shown how to use it according to the instructions for
use.