If highly alkaline cleaners are used, metal surfaces may be discol-
ored. However, this does not affect the integrity of the product.
2. Check the mechanical functionality of the product:
– Can the jaws be opened and closed?
– Can the handle be opened and closed?
– ERGO handle (20195-245): Can the upper part of the
handle be rotated freely relative to the lower part? Can
the locking knob be locked and loosened?
If damaged, do not use this product!
3. Oil all moveable parts of the handle and the joint area of the
insert using a physiologically acceptable oil (e.g. paraffin oil).
7.9

Packaging

The product can be sterilized while it is dismantled or assembled.
The cable must be disconnected from the product. The jaws must
be open.
The product must be protected against damage during steriliza-
tion.
1. If required, assemble the product (no cable). See page 24.
2. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/foil and/or in a ster-
ilization container.

7.10 Sterilization

Only sterilize products that have been cleaned and disinfected.
Erbe Elektromedizin recommends steam sterilization with the
method described below. If other sterilization methods are used,
Erbe Elektromedizin shall bear no responsibility.
32