OBJ_DOKU-180148-009.fm Seite 24 Donnerstag, 17. Januar 2019 10:12 10
3
Maximum electrical capacity
The maximum electrical capacity is given on the product in [V
4
Safety instructions
WARNING!
This product may be used only by trained medical staff, who
have been shown how to use it according to the instructions for
use.
Before every use check the insulation of the connecting cables,
the activation functions of the product and whether the product
has any physical damage.
Erbe Elektromedizin recommends performing an electrical con-
tinuity test before every use of this product.
Note: If the cable conductor is broken or the electrical continuity
of the cable is interrupted in any other manner it can cause arc-
ing in the patient return system or in the active circuit; this can
lead to the patient suffering burns or the outbreak of a fire.
Checks other than those described in these instructions for use
do not need to be carried out.
If damaged, do not use this product!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Protect this product from any form of mechanical damage! Do
not throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
24
].