Regular inspection and adjustments (Continued)
EN
Item
Litter
Mattress
Base
Accessories (optional)
Every six months or 12 hours
Inspect these items every six months or 12 hours, whichever comes first.
Item
Litter
Base
Wheels
Every 12 months or 24 hours
Inspect these items every 12 months or 24 hours, whichever comes first.
Item
Litter
1-34
Preventive maintenance
Inspect
All fasteners are secure
Backrest cylinder operates
Adjust pneumatic cylinder for full range of motion, if
required
Dead stop (6085-001-094) for wear, replace if required
No cracks or tears
All fasteners are secure
Plastic wear strips are intact on the front and rear leg
assemblies
All accessories operate (such as IV pole, head extension
and pillow, defibrillator platform, and Pedi-Mate®
restraint package)
Inspect
No bent, broken, or damaged components
No damage or tears on cot grips
Siderails operate and latch
Footrest operates
No bent, broken, or damaged components
Free of debris
Inspect
All welds are intact, not cracked, or broken
Warning labels present, legible
Fowler bumper and screws intact and secure
Perimeter bumpers are intact
Backrest operates
Extension lift handles operate
Siderails operate and latch
Knee gatch/Trendelenburg operates