6.3

Printing out measured values (PRINT/PRINT MEM button) (C.A 6543)

If you use a serial printer, choose the appropriate communication speed in the SET-UP menu, between
300...9600 bauds, then program the printer to the format run by the instrument (see § 6.1).
If you are using a parallel printer, you should set the signalling speed to "Parallel" when in SET-UP and
use the serial-to-parallel adapter that comes as an optional item (to plug in series the supplied cable +
adapter + Centronics printer cable).
There are two different printing modes possible :
- Instantaneous printing of measurement (PRINT)
- Printing of recorded data (PRINT memory)
!
If the data transmission to the printer is successful, the COM symbol flashes on the screen.
If a problem is experienced, the COM symbol remains permanently on the LCD screen without flashing.
6.3.1
Instantaneous printing of measurements (PRINT button)
Following a measurement or after accessing values in MR mode, use the PRINT function to print-off the
measurement results.
Pressing the button brings about the printing of:
- 1 set of measurements (U/R/DAR/PI/date/time) for normal tests,
- the R(t) values if the 'time-controlled test' has been used.
To stop printing, alter the position of the selector switch.
According to the function used, the following models are obtained:
I I I I I
Insulation measurement
CHAUVIN ARNOUX C.A 6543
Instrument number: 000 001
INSULATION RESISTANCE TEST
OBJECT : 01
Description : .................................................
.....................................................................
Date : ......................................... 31.03.1998
Start time: .......................................... 14 :55
Test duration: ....................... 15 min. 30 sec
Temperature : ............................. °C .........°F
Relative humidity : .................................... %
Test voltage : ..................................... 1000 V
Insulation resistance (R) : ........ 385 GOhm
DAR (R 1'/R 30") : ............................... 1.234
PI (R 10'/R 1") : ................................... 2.345
Comments: ..................................................
.....................................................................
Date of next test : .......................... /..../.......
After a 'time-controlled test' other results are printed (interval samples) :
Time
Resistance
00 : 30
35.94 GOhm
01 : 00
42.00 GOhm
01 : 30
43.50 GOhm
etc...
TEST : 01
(only printed in MR mode)
Voltage
1005 V
1005 V
1005 V
A space for the operator to fill in their signature
comes at the end of the printout.