Operating Environment
Battery Type
Output Power
Battery Operating Time (est.) Battery: External NiMH, charged in safe area. 1 load cell - 30-60 hours; 4
Battery Charging Time (est.)
Hazardous Area Approvals
64062816
|
09
|
05/2021
Specifications
Can be operated at temperatures ranging from −10° to 40° C (14° to 104° F)
at 10% to 95% relative humidity non-condensing.
Consists of eight NiMH battery cells in an encapsulated pack for a total power
of 8Ah.
Output voltage 7.5 to 10 VDC nominal @ 130mA minimum (170mA
minimum for IDNet) to 3 A maximum. A 3m (10 ft.) cable included in
64060627 kit. Cable not designed to be extended.
load cells - 25-35 hours
11 hours
United States
30538111 battery pack (FM19US0232)
Class I,II,III; Division 1; Groups A-G; T5
Class I Zone 1 IIC T5
ZONE 21 IIIC T93°C
-10°C ≤ Ta ≤ +40°C IP66
64060625 battery pack (FM19US0149)
Class I,II,III; Division 1; Groups A-G; T5
Class I,II,III; Division 2; Groups A-D, F,G; T5
Class I Zone 1 AEx ia IIC T5
-10°C≤Ta≤+40°C IP66
30538111 battery pack (FM19CA0215)
Canada
Class I,II,III; Division 1; Groups A-G; T5
Zone 1 - Ex ib IIC T5 Gb
Zone 21 - Ex ib IIIC T93°C Db
-10°C ≤ Ta ≤ +40°C IP66
64060625 battery pack (FM19CA0079)
Class I,II,III; Division 1; Groups A-G; T5
Class I,II,III; Division 2; Groups A-D, F,G; T5
-10°C≤Ta≤+40°C IP66
30538111 battery pack (FM19ATEX0216)
Europe (ATEX)
Zone 1 - II 2 G Ex ib IIC T5 Gb
Zone 21 - II 2 D Ex ib IIIC T93°C Db
-10°C ≤ Ta ≤ +40°C IP66
64060625 battery pack (FM07ATEX0043)
Zone 1 - II 2 G Ex ib IIC T5 Gb
Zone 21 - II 2 D Ex ibD 21 T93°C IP66 Db
-10°C ≤ Ta ≤ +40°C
30538111 battery pack (IECEx FMG 19.0046)
IECEx FMG
Zone 1 - Ex ib IIC T5 Gb
Zone 21 - Ex ib IIIC T93°C Db
-10°C ≤ Ta ≤ +40°C IP66
64060625 battery pack (IECEx FMG 12.0026)
Zone 1 - Ex ib IIC T5 Gb
Zone 21 - Ex ibD 21 T93°C Db
-10°C ≤ Ta ≤ +40°C IP66
METTLER TOLEDO NiMH Battery Pack Installation/Technical Manual