Hach AS950 Manual Del Usuario página 7

Ocultar thumbs Ver también para AS950:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 108
5.1.2 Configure the hardware settings
1. Push MENU.
2. Select Hardware Setup.
3. Select an option.
Option
Sampler
Configure
Sensors
AUX and I/O
Port
Liquid Detect
Rain and
RS485 Port
Thermal
(AWRS
sampler)
5.1.2.1 Configure the sensors
1. Push MENU.
2. Select Hardware Setup>Configure Sensors>[Select Sensor].
3. Select an option.
Note: Not all of the options that follow apply to all sensors.
Refer to
Table 4
for descriptions of all the pH sensor settings. Refer to
all the US900x sensor settings. Refer to
settings.
Option
Setup Wizard
Calibration
Basic Settings
Flow Settings
2
Calibrate the liquid detector only with the guidance of technical support.
Description
Site ID—Enters the name for the sampling location (16 characters). Rinses
—Sets the number of intake tube rinses before a sample is collected (0 to 3).
Sample Retries—Sets the number of sampling tries after a sampling failure
occurs (0 to 3).
Configures the attached sensors. Refer to
on page 7.
Type—Configures the AUX I/O port. Refer to
on page 10.
Enables or disables the liquid detector. Calibrates the liquid detector
Enabled (default)—Select enabled to use the liquid detector to sense liquid
in the tube, which is an important part of volume accuracy and repeatability.
Make sure to select enabled in applications where liquid level changes occur.
Disabled—Select disabled when the sample is collected from a pressurized
line or when the sampler is below the sample source, which can result in
trapped water.
Configures the Rain port for connection to the Hach tipping bucket rain
gauge or RS485 communications. Options: Rain or RS485.
When RS485 is selected, sets the Modbus address (1 to 255), the baud rate
(9600, 19200, 38400, 57600 or 115200) and the parity (None, Even or Odd).
Sets the temperature inside the AWRS refrigerated cabinet. Options: 2.00 to
10.00 °C (default: 4.00 °C). Calibrates the temperature inside the AWRS
refrigerated cabinet.
Table 6
Description
Configures all the sensor settings. Calibrates the sensor. Obey the
screen prompts to complete the procedures. Refer to
sensors
on page 20.
Calibrates the sensor. Refer to
Configures the basic sensor settings.
Configures the sensor settings that are used to calculate the flow.
Configure the sensors
Configure the AUX I/O port
for descriptions of all the AV9000 sensor
Calibrate the sensors
2
.
Table 5
for descriptions of
Calibrate the
on page 20.
English 7

Publicidad

Tabla de contenido
loading

Tabla de contenido