*The values for current and voltage are given as numbers, scaled
from 0 to 255, where 0 corresponds to 0 V and 255 to the full scale
voltage (130 V) or current (5 A).
When the Request is "G", the Answer will contain a string of 12
characters with the following parameters:
Tab. 7
BYTES
1
2-3
4
5-8
9
10
11
12
When the configuration parameters have to be changed, send a
Request string with "H" characters of the following type:
"H" + .... DATA .... + CRC
The CRC is calculated on the string consisting of H + DATA.
The following parameters can be changed:
Tab. 8
BYTES
1
2
3-6
7
8
9
10
Turbo-V 700HT Controller User Manual / 87-900-877-01 (F)
MEANING
Speed adjust value (in KRPM)
Pump cycles number (integer coded in 2 bytes)
Speed threshold
Run up time in seconds (long coded in 4 bytes)
Deat time (0 = NO 1 = YES)
Soft Start mode (0 = NO 1=YES)
Water cooling (0 = NO 1=YES)
CRC
MEANING
Speed adjust value (in KRPM)
Speed threshold
Run up time in seconds (long coded in 4 bytes)
Deat time (0 = NO 1 = YES)
Soft Start mode (0 = NO I=YES)
Water cooling (0 = NO I=YES)
CRC
Technical Information
Optional Serial Port