Figure 21
where:
Turbo-V 700HT Controller User Manual / 87-900-877-01 (F)
Press the CURRENT push-button and the display shows:
=
0
.
0
.
.
I = is the DC current drawn by the pump range (0.00 to 10.00
Ampere)
P= is the DC power drawn by the pump (range 0 to 999 Watt)
KRPM = is the theoretical rotational speed of the pump as a
function of the controller output frequency (range 3 to 42 KRPM)
°C = is the temperature of the outer ring of the upper bearing
(range 00 to 99°C)
X = during operation a selected set point condition (1 or 2
contrast inverted) appears when the programmed threshold
speed value is not reached.
0