Fuse holder and voltage changer assembly
The following figure shows the location of this assembly.
Figure 6
Proceed as follows to replace one or both fuses:
1
2
Use only T-type fuses of the following characteristics:
Turbo-V 700HT Controller User Manual / 87-900-877-01 (F)
Rear panel
Remove the assembly by levering in position A with a small
screwdriver.
Replace the fuse.
100/120 Vac 6.3 A
220/240 Vac 3.15 A
Technical Information
Fuse holder and voltage changer assembly
17