KRPM
R1
ON
OFF
R2
ON
OFF
R1
ON
OFF
Figure 32
Turbo-V 700HT Controller User Manual / 87-900-877-01 (F)
RUN-UP TIME
Cycle diagram
Technical Information
case
a = pump speed > threshold value
programmed speed threshold value
b = pump speed < threshold value after run-up time
c = pump speed < threshold value before run-up time
t
R1 operation
a
b,c
t
b,c
R2 operation
delay even after YES
a
t
b,c
R2 operation
delay even after NO
a
t
17
Use
283/304