Medela Pump In Style Instrucciones De Uso página 34

Extractor de leche eléctrico doble
Ocultar thumbs Ver también para Pump In Style:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

12. EMC technical description (cont.)
Guidance and manufacturer's declaration – electromagnetic immunity
34
This breast pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the breast pump should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
test level
3 V
0,15 MHz – 80 MHz
6 V in ISM and
amateur radio bands
between 0,15 MHz
and 80 MHz
80% AM at 1 kHz
10 V/m (minimum)
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance
Electromagnetic environment –
level
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the breast pump, including cables,
than the recommended separation
distance calculated from the equation
6V
applicable to the frequency of the
transmitter.
Recommend separation distance
d = 1,2√P
d = 1,2√P 80 MHz - 800 MHz
d = 2,3√P 800 MHz - 2.7 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
10 V/m
recommended separation distance in
(minimum)
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol.
should
a
b