5.2
Processing limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
5.3
Dismantling
1. Disconnect the adapter from the FiAPC probe. The FiAPC
probe cannot be processed.
5.4
Required aids
Processing step
Items Needed
Precleaning
Soft plastic brush/soft cloth
50 ml disposable syringe with male
Luer-Lock connection
Female-to-female luer-lock adapter
(=irrigation adapter)
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
50 ml disposable syringe with male
Luer-Lock connection (4 ea.)
Female-to-female luer-lock adapter
(=irrigation adapter)
Compressed air gun (2 bar maximum)
Machine cleaning/
Female-to-female luer-lock adapter
disinfection
(=irrigation adapter)
27