7.3
Preparations at the place of use
►
Remove all attached components from the product (tool an accesso-
ries).
►
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
►
Place the dry product in a sealed waste container and forward it on for
cleaning and disinfection within 6 hours.
7.4
Preparation before cleaning
►
Prior to the first mechanical cleaning/disinfection process, mount the
ECCOS fixations in a suitable tray (e.g. GB243800).
►
Insert the products in the correct position into the ECCOS fixations, see
Fig. E.
7.5
Cleaning/disinfection
7.5.1
Product-specific safety instructions for the reprocessing pro-
cedure
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
►
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– are approved for plastic material and high-
grade steel,
– do not attack softeners (e.g. in silicone).
►
Do not use cleaning agents that contain ace-
tone.
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum temperature of
60 °C during chemical cleaning and/or disinfec-
tion.
►
Do not exceed the maximum temperature of
96 °C during thermal disinfection with FD
water.
►
Dry the product for at least 10 minutes at a
maximum of 120 °C.
Damage or destruction of the batteries due to pro-
cessing!
►
Protect battery from moisture.
CAUTION
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.
9