23
2. Set the supplemental gas flow to 15 liters per minute for the adult and child models; and to 10 l/m for
the infant model.
3. Cycle the Resuscitator through several ventilations. The test lung should inflate during inspiration and
deflate during exhalation. Check for leakage at all joints and connections. Ensure that the Resuscitator
refills promptly and properly and that all valves are operating correctly. If not, repeat the tests above to
find where the problem lies.
Specification & Performance
Storage Temperature: -40°C (-10°F) to 60°C (140°F)
Operating Temperature: -18°C (0°F) to 50°C (122°F)
Materials
Silicone rubber
silicone bag
duckbill valve
flapper valves
relief valve seal
infant mask
child mask
adult mask bladder
mask grommet
mask retainers
Epichlorohydrin
O-ring (bag neck)
Stainless Steel
pressure relief valve spring
Connectors
patient port
SILICONE BAG INLET
reservoir valve
supplemental gas inlet
ENGLISH
Polycarbonate
nonrebreathing valve housing
bag intake valve housing
reservoir valve housing
reservoir bag connector
adult mask shell
pressure relief valve housing
pressure relief valve stem
Polyvinylchioride
oxygen reservoir bag
15 mm I.D.(F) / 22 mm O.D.(M)
23 mm I.D.(M)
23 mm I.D.(F) (to silicone bag inlet)
25 mm O.D.(M) (to oxygen reservoir)
6mm O.D.(M)