WIKA ACS-10 Manual De Instrucciones página 80

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

5. Commissioning, operation
EN
Furthermore, the temperature of the test item must be checked in the field "Gauge
temperature" (1).
If the temperature deviates considerably from the temperature probe on the hose pack,
a manual entry can also be made. However, this is not recommended as the correct
temperature measurement cannot be ensured.
In general, sufficient time and, if possible, the same ambient conditions should be
ensured during temperature control.
Confirm the entries with the "Confirm" button (2).
The actual testing process starts.
80
For a recalibration to be as accurate as possible, it must be ensured that
the test item has the same temperature as the reference thermometer of
the calibration system. For this, either sufficient time must be planned for
temperature equalisation or the temperature must be entered manually.
Deviations may occur if the ACS-10 is transported in an air-conditioned
vehicle, for example, and a measurement is carried out immediately
afterwards.
WIKA Betriebsanleitung Kalibriersystem, Typ ACS-10

Publicidad

Tabla de contenido
loading

Tabla de contenido