Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
0 % U
for 0.5 cycle
T
at 0°, 45°, 90°, 135°,
180°, 225°, 270°,
and 315°
0 % U
for 1 cycle
T
at 0°
70 % U
for 25
T
cycles at 50 Hz
at 0° for 30 cycles
at 60 Hz at 0°
0 % U
for 250
T
cycles at 50 Hz at
0° for 300 cycles
at 60 Hz at 0°
30 A/m
0 % U
for 0.5 cycle
T
at 0°, 45°, 90°, 135°,
180°, 225°, 270°,
and 315°
0 % U
for 1 cycle
T
at 0°
70 % U
for 25
T
cycles at 50 Hz
at 0° for 30 cycles
at 60 Hz at 0°
0 % U
for 250
T
cycles at 50 Hz at
0° for 300 cycles
at 60 Hz at 0°
30 A/m
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
breast pump requires
continued operation
during power mains
interruptions, it is
recommended that
the breast pump be
powered from an
uninterruptible power
supply or a battery.
It may be necessary
to position the breast
pump further from
sources of power
frequency magnetic
fields or to install
magnetic shielding.
The power frequency
magnetic field should
be measured in the
intended installation
location to assure that
it is sufficiently low.
39