NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
Wi-Box™ AO transmitter is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Table 10. Guidance and manufacturer's declaration – Electromagnetic immunity
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5
GHz
*Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the Wi-Box AO transmitter is used exceeds the applicable RF compliance level above, the device should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the device.
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
1
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of this device,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
3 Vrms
Recommended separation distance
d=1,2√P
3 V/m
d=1,2√P 80 MHz to 800 MHz
d=2,3√P 800 MHz to 2,5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey*, should
be less than the compliance level in each frequency
range
.
1
Interference may occur in the vicinity of equipment
marked with the following symbol.