Figure 2. Connecting AO transducer to Wi-Box AO transmitter if a multi-channel cable adapter is in
use
Make sure that the blood pressure channel is powered.
1.
To verify correct pressure, apply a known pressure on the blood pressure transducer using a
2.
blood pressure simulator.
If the green READY light is lit and the AO pressure is correctly displayed on the hemodynamic
recording system, the Wi-Box AO transmitter is ready to use.
If the red NOT READY light is lit or the AO pressure is incorrectly displayed on the hemodynamic
recording system, the Wi-Box™ external power supply needs to be connected to provide adequate
power to the unit (see the Wi-Box External Power Supply section below).
If the red NOT READY light is blinking, internal fail has occurred. Contact your St. Jude Medical
representative.
Wi-Box™ External Power Supply (for Low Power Hemodynamic
Recording Systems)
Connect the Wi-Box external power supply cable to the PSU connector on the Wi-Box™ unit.
1.
Connect the Wi-Box external power supply to a free wall socket using the appropriate wall
2.
socket adapter.
The white light on the Wi-Box unit shall now be lit, indicating that the device is powered.
To verify correct pressure, apply a known pressure on the blood pressure transducer using a
3.
blood pressure simulator.
If the green READY light is lit and the AO pressure is correctly displayed on the hemodynamic
recording system, the Wi-Box AO transmitter is ready to use.
If the red NOT READY light still is lit, an internal error has occurred. Try to reconnect the
Wi-Box external power supply. If the problem remains contact your St. Jude Medical
representative. If the AO pressure is incorrectly displayed on the hemodynamic recording
system, do not use the Wi-Box AO transmitter. Contact a St. Jude Medical representative.
CAUTION
 The only approved power supply to connect is
C12785/FW7333SM.
 The power supply shall be mounted at a location where it is
protected from fluids.
 Mains power is disconnected by removing the Wi-Box™
external power supply from the wall socket. Place the Wi-Box
external power supply such that it easily can be removed from
the wall socket.
4
1. AO transducer
2. Wi-Box AO transmitter
3. Hemodynamic recording
system
4. Multi-channel cable
adapter
5. Single-channel cable

Publicidad