P1.2: READY indicator: (rdy)
Meter is defaulted as "ON". Icon displays when measurement gets
stable. Users can turn it off for faster response. Press UP /DOWN to
switch "on" and "off". Press ENTER to confirm and enter P1.3.
P1.3: ATC or Manual TC: (Atc)
Meter is defaulted as ATC on. Press UP/DOWN to switch on and off.
Press ENTER to confirm and return to P1.0.
P2.0: Unit : (Unt)
P2.1: Select °C or °F:(t)
Select P2.0 and press ENTER to P2.1. Default unit oC is blinking on LCD.
Press UP/DOWN to switch oC and oF. Press ENTER to confirm and enter
P2.2.
P2.2: Select ppm or mg/L: (tdS)
The default TDS unit "ppm" blinks on LCD. Press UP/ DOWN to switch
between ppm and mg/l, Press ENTER to confirm and return to P2.0.
P3.0: temperature parameters: (t)
P3.1: Temperature coefficient: (tCo)
The temp. coefficient (expressed in percent per °C) is the changed ratio
of conductivity per degree of temp..The adjustable range is 0.0 per °C
to 10.00 % per °C. The default is 2.10% per °C. 0.0% has no effect on
temp. So the displayed value is the same as actual temperature. Select
P3.0 and press ENTER to enter P3.1.Press ENTER again and Temperature
Coefficient flashes on LCD (Fig.4). Press UP/DOWN to change the value
from 0.0 to 10.0. Press ENTER to confirm and enter P3.2.
Fig. 4
6