Descargar Imprimir esta página

Keysight U1251B Guia Del Usuario página 158

Ocultar thumbs Ver también para U1251B:

Publicidad

6
Performance Tests and Calibration
Table 6-2
Verification Test (continued)
Step
Test Function
4
Turn the rotary switch to
position
(for model U1252B),
to
position
(for model U1252B)
5
Press
the
V mode
6
Turn the rotary switch to the
position
156
Range
5 V
50 V
500 V
1000 V
5 V
button to go to
[a]
50 V
500 V
1000 V
50 mV
500 mV
1000 mV
5520A Output
5 V
50 V
500 V
1000 V
5 V,1 kHz
5 V, 10 kHz
4.5 V, 20 kHz
4.5 V, 100 kHz
50 V, 1 kHz
50 V, 10 kHz
45 V, 20 kHz
45 V, 100 kHz
500 V, 1 kHz
1000 V, 1 kHz
50 mV
500 mV
– 500 mV
1000 mV
– 1000 mV
Keysight U1251B and U1252B User's and Service Guide
Error from nominal 1 year
U1251B
U1252B
± 2 mV
± 1.75 mV
± 20 mV
± 17.5 mV
± 200 mV
± 200 mV
± 800 mV
± 800 mV
N/A
± 22.5 mV
N/A
± 79.0 mV
N/A
± 169.5 mV
N/A
± 169.5 mV
N/A
± 225 mV
N/A
± 790 mV
N/A
± 1.695 V
N/A
± 1.695 V
N/A
± 2.25 V
N/A
± 8.0 V
[b]
± 75 V
± 75 V
± 0.2 mV
± 0.175 mV
± 0.2 mV
± 0.175 mV
± 0.8 mV
± 0.75 mV
± 0.8 mV
± 0.75 mV
[b]