2
Making Measurements
Measuring Temperature
Do not bend the thermocouple leads at sharp angles. Repeated bending
CAUTION
over a period of time may break the leads.
The bead type thermocouple probe is suitable for making temperature
measurements between –20 °C to 200 °C in PTFE compatible environments.
Do not use the bead-type thermocouple probe beyond the recommended
operating temperature range. Do not immerse this thermocouple probe in liquids.
For best results, use a thermocouple probe designed for each application — an
immersion probe for liquid or gel and an air probe for air measurements.
Set up the multimeter to measure temperature as shown in
observe the following steps:
1 Press
2 Connect the miniture thermal probe into the non-compensation transfer
3 Connect the thermal probe with the adapter into the meter input terminals as
4 For best performance, place the meter in the operating environment for at
5 Clean the measurement surface and make sure the probe is securely touching
6 When measuring above the ambient temperature, move the thermocouple
7 When measuring below ambient temperature, move the thermocouple along
70
to select the temperature measurement.
adapter as shown in
shown in
Figure
2-14.
least one hour to stabilize the unit to environment temperatures.
the surface. Remember to disable the applied power.
along the surface until you get the highest temperature reading.
the surface until you get the lowest temperature reading.
Figure
2-13.
Keysight U1251B and U1252B User's and Service Guide
Figure 2-15
or