Table 6-4
Adjustment table (continued)
Step
Test Function
9
Press
[b]
mode
10
Turn the rotary swich to the
position
11
Press
μA mode
Keysight U1251B and U1252B User's and Service Guide
Cal Range
10 nF
100 nF
1000 nF
10 μF
100 μF
1000 μF
10 mF
N/A
button to go to
OPEN
500 μA
5000 μA
500 μA
button to go to
5000 μA
Performance Tests and Calibration
Input
U1251B
3 nF
03.000 nF
10 nF
10.000 nF
10 nF
010.00 nF
100 nF
100.00 nF
100 nF
0100.0 nF
1000 nF
1000.0 nF
10 μF
10.000 μF
100 μF
100.00 μF
1000 μF
1000.0 μF
10 mF
10.000 mF
0 ºC
0000.0 ºC
Input terminal open
oPEn
(remove any test
leads and Short
Plugs from the
input terminal)
300 μA
300.00 μA
3000 μA
3000.0 μA
30 μA,1 kHz
030.00 μA
300 μA,1 kHz
300.00 uA
300 μA,1 kHz
0300.0 μA
3000 μA, 1 kHz
3000.0 μA
6
Cal Item
U1252B
03.000 nF
10.000 nF
010.00 nF
100.00 nF
0100.0 nF
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
0000.0 ºC
oPEn
300.00 μA
3000.0 μA
030.00 μA
300.00 μA
0300.0 μA
3000.0 μA
173