Descargar Imprimir esta página

Keysight U1251B Guia Del Usuario página 173

Ocultar thumbs Ver también para U1251B:

Publicidad

Table 6-4
Adjustment table (continued)
Step
Test Function
3
Press
the
V mode
4
Turn the rotary switch to the
position
5
Press
the
mV mode
Keysight U1251B and U1252B User's and Service Guide
Cal Range
5 V
button to go to
50 V
500 V
1000 V
Short
50 mV
500 mV
1000 mV
50 mV
button to go to
500 mV
Performance Tests and Calibration
Input
U1251B
0.3 V, 1 kHz
N/A
3 V, 1 kHz
N/A
3 V, 20 kHz *
N/A
3 V, 1 kHz
N/A
30 V, 1 kHz
N/A
30 V, 20 kHz *
N/A
30 V, 1 kHz
N/A
300 V, 1 kHz
N/A
300 V, 20 kHz *
N/A
30 V, 1 kHz
N/A
300 V, 1 kHz
N/A
300 V, 20 kHz *
N/A
Dual banana plug
SHort
with copper wire
short between 2
terminals
30 mV
30.000 mV
300 mV
300.00 mV
1000 mV
1000.0 mV
3 mV, 1 kHz
03.000 mV
30 mV, 1 kHz
30.000 mV
30 mV, 20 kHz *
30.000 mV
30 mV, 1 kHz
030.00 mV
300 mV, 1 kHz
300.00 mV
6
Cal Item
U1252B
0.3000 V
3.0000 V
3.0000 V
03.000 V
30.000 V
30.000 V
030.00 V
300.00 V
300.00 V
0030.0 V
0300.0 V
0300.0 V
SHort
30.000 mV
300.00 mV
1000.0 mV
03.000 mV
30.000 mV
30.000 mV
030.00 mV
300.00 mV
171

Publicidad

loading

Este manual también es adecuado para:

U1252b